Analyzer BAS-100 TS is a TOUCH SCREEN Microprocessor Biochemical Analyzer, which measures the biological indicators, more than 50 elements.
The BAS-100 TS is open to being used with all brands of biochemistry reagents.
Single chip machine, WindowsCE operating system, does not require Windows XP or Windows 7.
THIS BIOCHEMISTRY ANALYZER CAN TEST SAMPLES IN TWO WAYS,
BY SUCTION PUMP AND BY DISPOSABLE CUVETTE
Features:
Optics: 7 filters, 340, 405, 450, 505, 546, 578, 630.
Thermostatic Control: Flow Cell Temperature: 37C Peltier Elements.
Flow through cell: 32ul
Suction Pump System
Display:7.0 inch TFT Color touch screen
Precision: +-0.2C
Continuous Memory: 200 assay program,200000 results
Printer: Built-in thermal printer, 57mm. wide paper
Power Supply: AC85-260; 50Hz or 60Hz
Sample blood serum
Volume: Reagent plus sample serum 500 ul
Reagent will give ratio
The suction volume is 500 ul
Halogen lifetime is 1500 hours
Calibration is leaner
Photometric resolution is 0.0002A
- Photometric scanning of Spectrum, Analysis in multiple longitude of wave and kinetics, endpoint, fixed time, kinetic. Blank of water or blank of reagent method can be linear or nonlinear, according to the type of reagent you are using.
- Diagnosis of Operation
- Automatic Calibration
- System of aspiration of sample (Pale of flow, Aspiration System, Faslk of Residuals, Tubes, Connectors, Etc)
- Power Cable Included
- Works with WindowsCE operating system. Single chip machine, does not require Windows XP or Windows 7.
- Enzyme Kinetic GPT GOT tests.
Assay method: |
Absorbance, concentration, factor, 1 point end point, 2 point end (Sample blank), fixed time, kinetic, and dichromatic, immunoturbidimetric. Coagulation (optional). |
Wave Length: |
7 filters, 340, 405, 450, 505, 546, 578, 630 as standard configuration. (5 years warranty for each filter) |
Light source: |
6V 10W Long-life Tungsten halogen lamp. Service life 2000 hours.
Light has auto shutdown function that will extend usage period.
Auto-Sleep function. |
Display: |
7.0 inch TFT Color touch screen,260000 Pixel,800x480 resolution; real time to show response curve |
Built-in Thermal Printer: |
Built-in Thermal Printer, 57mm printer paper
Paper alarm, transparent cover; easy installation |
Incubation temperature: |
37℃ and room temperature are standby; temperature accuracy±0.1℃,
by means of Peltier elements |
Flow cell capacity and optical path: |
32ul, 10mm |
Resolution: |
0.0001 Abs |
Photometric Range: |
0 – 3.5 Abs |
Absorbance range: |
0.0000-2.5000(10mm flowcell)
0.0000-3.5000(6mm flowcell) |
Spectral Band: |
< 8 nm
|
Vacuum system programmable:
|
200 – 5000 ul at least |
Carry over: |
<1%; |
Stability: |
<0.005Abs/hour |
Absorbance Precision: |
0.0001Abs; |
Absorbance drift: |
<0.005Abs/hour |
Absorbance repeatability: |
CV≤1% |
QC analysis: |
L-J QC chart , Westguard Multi-regulation analysis. |
Software: |
Window CE 6.0 operation system,high-capacity storage; edit 200 assay program,200000 results
Unlimited storage by connecting to computer.
Single chip machine, does not require Windows XP or Windows 7. |
Operation Manual: |
Available in 4 different languages |
CPU: |
ARM Cortex-A8, 720MHz |
Memory: |
DDR2 SDRAM: 1G bit
Flash: 256 MB
Over 200 programs, 200,000 results can be stored |
USB Port: |
USB Mouse; USB keypad; USB Printer; 4 USB ports for Keyboard, Mouse, Printer, U Disk, Bar Code Reader. 1 USB Slave for PC connections (supports PC software control function).
1 Ethernet port.
Supports all HP series inkjet printers; Computer control software,LIS system available. External scanner is available. Support remote diagnostics |
Operational Environment: |
Temperature: 10℃~37℃. Humidity: ≤85%. |
Power supply: |
Wide power supply: 100-240V 50Hz or 60Hz. |
Dimensions: |
400mm (L) x 300mm (W) 160mm (H) |
Weight: |
17 lbs.. |